HSSC-A6series

Specification

Dimension

50kPa~3MPa

HSSC-A6_High temperature type

50kPa~3MPa

HSSC-A6_DIN connector type

Digital pressure meter with large high-brightness LED(14.2mm) thanks to 2-wire transmission system.
Constructed with stainless steel in view of environmental resistance.
Safe and reliable construction with break-proof diaphragm.
Safe and reliable construction with no sealed liquid.

Measured mediumAlcohols, milk beverage, seasoning, pure water, chemical or other media that will not corrode SUS316L.
Pressure typePositive, compound, and negative pressures can be measured.
Pressure rangeCan be specified within the range of 50 kPa to 3 MPa (Calibration:25kPa~3MPa) of rated capacity.
Specify more than 50% of rated capacity as the calibration range.
Output type4~20mA (2 wire system)

Specification 1 (Sensor assembly)

Pressure Port Type/SizeISO Ferrule, VARIVENT®, ISO Cap nut, Screw mount, Other special orders
Rated Capacity(R.C.)50, 100, 200, 300, 500, 1000kPa, 3MPa
MaterialSensor: SUS316L
Finish(Sensor Part)Electro-Chemical Buffing
Safe Overload Rating400%R.C. (Rated capacity 3MPa: 150%R.C.)
Destruction Point1000%R.C. (Rated capacity 3MPa: 300%R.C.)

Specification 2 (Amplifier + Sensor assembly)

Accuracy (*1)
(Non Linearity, hysteresis and repeatability included.)
±0.3%R.C. (Rating: 300, 500, 1000kPa, 3MPa)
±0.35%R.C. (Rating: 50, 100, 200kPa)
Compensated Temperature Range (Wetted part)High temperature: -5~150?C
Safe Temperature Range (Wetted part)High temperature: -5~150?C
Temperature Effect on Zero Balance±0.35%R.C./10?C (Rating: 300, 500, 1000kPa, 3MPa)
±0.3%R.C./10?C (Rating: 50, 100, 200kPa)
Temperature Effect on Load±0.35%R.C./10?C (Rating: 300, 500, 1000kPa, 3MPa)
±0.3%R.C./10?C (Rating: 50, 100, 200kPa)
Ambient Temperature and HumidityTemperature:0~50?C(No freezing), Humidity:0~90%RH(No condensation)
MaterialSUS304(Case material), SUS304(Radiating fin)
IP RatingEquivalent to IP65
WeightApprox. 700g
AccessoriesOperation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy)
1 of hexagon nut(Cap nut only)
  • (*1) For the ferrule type, zero adjustment is to be made after the clamp is tightened.

Model Selection

Model Selection

SelectionCheckModelSpecification
1.ModelHSSCHigh temperature type: Temperature range of wetted parts -5~150?C Stainless steel case
2.Display/Analog outputA63 ? digit Red LED Character height: Approx.14.2mm
Display conversion rate: Fixed at approx. 2.5 times per second
4~20mA(2 wire system) Response:50Hz Safe load resistance:400Ω or less
3.Sensor ConnectionASSensor directly connected on bottom(Flangeless)
4.Display angle
Not availableDisplay angle: Front side, upward

(3)Display angle: Front side, rightward

(6)Display angle: Front side, downward

(9)Display angle: Front side, leftward
5.Display/Calibration--Example of calibration range

PressurizationRated capacity
At 1000kPa1000kPa
(0~1000)kPa
Specify the values and units of display/calibration ranges

Negative pressureRated capacity
At 100kPa-100kPa
(0~-100)kPa
In case of negative pressure, put the calibration range value in parentheses.(*2)

Compound pressureRated capacity
At 500kPa-100~500kPa
(-100~500)kPa
Put the output calibration range value and unit in parentheses.(*2)(*3)
6.Rated Capacity(R.C.)
00550kPa

010100kPa

020200kPa(*3)

021200kPa(*3)
(Specify this model at calibration for compound pressure if "focus on negative pressure" has been selected.)

030300kPa(*3)

031300kPa(*3)
(Specify this model at calibration for compound pressure if "focus on negative pressure" has been selected.)

050500kPa

1001000kPa

3003MPa[Make to order]
7.Pressure Port Type/Size
F1Ferrule 1.5S

F2Ferrule 2S

F3Ferrule 2.5S[Make to order]

F4Ferrule 3S[Make to order]

F6Ferrule 4S[Make to order]

V50AVARIVENT® DN50[Make to order]

V65AVARIVENT® DN65[Make to order]

V80AVARIVENT® DN80[Make to order]

V100AVARIVENT® DN100[Make to order]

R1Cap nut 1.5S[Make to order] Hexagon nut included

R2Cap nut 2S[Make to order] Hexagon nut included

M0Screw mount type simple substance Dedicated O-ring included (×1)

M1JIS20K50A With flange, Dedicated O-ring included (×1)

M2JIS20K80A With flange, Dedicated O-ring included (×1)
8.Pressure Port Material
BlankSUS316L, Electro-Chemical Buffing(Ry0.7)
⑥Supports all rated capacities.
⑦Supports all pressure port type/size.

HS1Hastelloy C276, Electro-Chemical Buffing(Ry0.7)
⑥Rated capacitor can be selected from among 300, 500 and 1000kPa.
⑦Pressure port type/size can be selected from among F1, F2, F3 and M0.

J4LSUS329J4L, Buffing(#400)
⑥Rated capacitor can be selected from among 300, 500 and 1000kPa.
⑦Pressure port type/size can be selected from among F1, F2, F3 and M0.
9.Limitation
H400%R.C.

Blank150%R.C.(3MPa only)
10.Cable Connection
DDIN connector type

BlankPigtail 2m
11.Power4DC24V(±10%)
  • (*2) Basically the calibration range is within the rated capacity, but other requirements are possible, as long as the range is 50% or more than the rating of the sensor used.
    Specify the display range between -200 and 200 digits for the 4 mA side and between 100 and 1999 digits for the 20 mA side.

  • (*3) In case of calibration for compound pressure, calibration is performed only for the positive pressure, and only measurement data are given for the negative pressure. In this case, acceptance criterion is within ±1%R.C. When focusing on the negative pressure, it is also possible to perform calibration for the negative pressure and to provide only measurement data (within ±1%R.C.) according to the request from the customer.
    Regarding calibration, specify 020, 030 (focus on positive pressure) or 021, 031 (focus on negative pressure) in ⑥ above when placing an order.